Skip to Content To homepage

qX2x2/Cary300

Dual, air-cooled, Peltier-control for the Cary 100 and Cary 300 Spectrophotometers
Product Overview

The qX2x2/Cary300 is a pair of cuvette holders providing independent Peltier temperature control for both the sample and reference cuvettes in Cary 100 and 300 UV-Vis-NIR Spectrophotometers. The qX2x2/Cary300 replaces the sample tray of the Cary 300. It is mounted on a base plate and has utilities brought to the front panel for convenient access. Use the qX2x2/Cary300 to set and maintain precise sample temperatures and stir at steady reproducible rates, while bathing the cuvettes in a controlled gaseous environment. Although dependent on rapid, precise Peltier technology, no circulating water is required for normal operation from 0 °C to 110 °C. Purchase the low temperature option, consisting of a windowed vest and capability for liquid cooling, and operate the qX2x2/Cary300 from 0 °C down to -30 °C. Purchase the extended temperature option to extend operation above 110 °C to 150 °C. Control the qX2x2/Cary300, using the TC 1B/Dual Temperature Controller included in the purchase price. For additional control and temperature monitoring, use Quantum Northwest’s computer control software, T-App, available for purchase. Alternatively, use an ADL program for temperature ramping with WinUV ADL Shell software.

Product Features

Rapid, precise Peltier temperature control
Precise stepper-motor-driven magnetic stirring
Dry gas purge for work at low temperature
Lid to improve temperature stability
Threaded inserts to mount windows or filters
Optical slit holders to control illuminated volume
Cuvette lifter for removal of cuvette
Probe input for sample temperature
Complete sample tray for mounting
NIST-traceable temperature calibration
ADL program for temperature ramping

  • Technical Specifications
    Number of cuvettes controlled
    2
    Typical use
    Absorbance (sample and reference)
    Standard cuvette size (outside)
    12.5 mm x 12.5 mm
    Optical ports per cuvette
    2
    Optical port dimensions
    12 mm high x 10 mm wide
    Cuvette z-height
    15 mm
    Windowed vest available
    yes
    Factory set temperature range
    -40 °C to +110 °C
    Temperature range easily achievable
    -5 °C to +110 °C
    Range with low temperature option
    -35 °C to +110 °C
    Range with extended temperature option
    -35 °C to +150 °C
    Temperature precision
    ±0.02 °C
    Thermistor probes accepted
    Probe/QNW3
    Magnetic stirring motor type
    Stepper under each cuvette
    Magnetic stirring speed range
    1-2500 rpm
    Default stirring speed
    1200 rpm
    Stirring speed best performance
    60-1800 rpm
  • Basic Package

    qX2x2 – Pair of temperature-controlled cuvette holders
    TC 1B/Dual – Dual-Channel Temperature Controller
    Base plates, posts and front panel
    NIST-traceable calibration and performance data
    ADL program for temperature ramping
    Magnetic stir bar, tubing for dry gas and cables

  • Accessories
Other Products to Consider
Back to top