Skip to Content To homepage

t2 Sport/Cary300

Peltier-driven temperature control for the Cary 100 and Cary 300 Spectrophotometers
Product Overview

The t2 Sport/Cary300 consists of a pair of cuvette holders, an inexpensive, compact, temperature-controlled cuvette holder in the forward sample position, and an uncontrolled holder for the reference cuvette in the rear. Use the t2 Sport/Cary300 for temperature ranges of about -15 °C to 110 °C. It is ideal for most biological studies, either to observe absorbance spectra at precisely maintained temperatures, or for temperature ramping, to observe how absorbance changes with temperature. Operate the TC 1/Single Temperature Controller using the menu button or control via program T-App. Use an ADL program for temperature ramping with WinUV ADL Shell software. The t2 Sport/Cary300 is mounted on a base plate and has utilities brought to the front panel for convenient access.

Product Features

Rapid, precise Peltier temperature control
Variable speed magnetic stirring
Dry gas purge for work at low temperature
Mounting base and light-tight front panel
Cuvette lifter for removal of cuvette
Probe input for sample temperature
NIST-traceable temperature calibration
ADL program for temperature ramping

  • Technical Specifications
    Number of cuvettes controlled
    1
    Typical use
    Absorbance
    Standard cuvette size (outside)
    12.5 mm x 12.5 mm
    Optical ports per cuvette
    2
    Optical port dimensions
    12 mm high x 10 mm wide
    Cuvette z-height
    8.5 mm or 15 mm
    Windowed jacket available
    no
    Factory set temperature range
    -40 °C to +110 °C
    Temperature range easily achievable
    -15 °C to +110 °C
    Temperature precision
    ±0.02 °C
    Temperature accuracy
    Better than ±0.15 °C from 0 ⁰C to +80 ⁰C
    Accuracy full range
    Better than ±0.3 ⁰C from ‐15 ⁰C to +110 ⁰C
    Magnetic stirring motor type
    Stepper
    Magnetic stirring speed range
    1-2500 rpm
    Default stirring speed
    1200 rpm
    Stirring speed best performance
    60-1800 rpm
  • Basic Package

    t2 Sport – Temperature-controlled cuvette holder
    TC 1/Single – Temperature Controller
    Base plate, post (standard z-height = 15 mm) and front panel
    Reference cuvette holder without temperature control
    BATH 10 – Circulator
    NIST-traceable calibration and performance data
    ADL program for temperature ramping
    Accessories – Magnetic stir bar, tubing and cables

  • Available Cuvettes

    Test

  • Accessories
Other Products to Consider
Back to top