qMC 16
Discover the qMC 16: a temperature-controlled microcentrifuge tube holder designed for biological applications. It offers precise temperature control from -45°C to 80°C. Powered by a thermoelectric system, the qMC 16 rapidly heats or cools samples and maintains target temperatures more efficiently than traditional systems.
BiochemistryThe qMC 16 is a high-performance, temperature-controlled sample holder designed for up to 16 microcentrifuge tubes (1.5/2.0 mL), bringing Quantum Northwest’s expertise in precise temperature control to molecular biology and biochemistry. This compact unit ensures exceptional thermal stability, maintaining sample temperatures with minimal fluctuation.
Operating between -45 °C and 80 °C, the qMC 16 features one of Quantum Northwest’s most advanced thermoelectric temperature control systems. Its design incorporates key elements from the Brr series, optimizing low-temperature performance. A heat exchanger with circulating coolant ensures precise thermal regulation, while inert gas purging at lower temperatures helps maintain optical clarity and improve insulation. Though capable of reaching higher temperatures, the unit is limited to 80 °C based on industry recommendations to prevent tube deformation. Like all Quantum Northwest sample holders, the qMC 16 provides unsurpassed temperature control, ensuring stable hold temperatures within a fraction of a degree.
Control is straightforward using the TC-1 Temperature Controller, included with purchase. Users can set temperatures, and ramp rates via the TC-1’s front panel or through T-App, Quantum Northwest’s optional control software, which provides features like real-time plotting, scripting, and direct sample temperature measurement when paired with the optional Probe/10K sensor. Even without T-App, the TC-1 display provides a direct readout of the sample temperature when using the Probe/10K.
The qMC 16 mounts easily to both imperial and metric optical breadboards. If the standard model does not fully meet your requirements, contact us to explore customization options.
With its exceptional temperature control, compact design, and compatibility with standard microcentrifuge tubes, the qMC 16 is an ideal solution for temperature-sensitive experiments in molecular biology, biochemistry, and life sciences. Whether you need stable-temperature incubation or precise temperature cycling, the qMC 16 ensures reproducible results in a versatile, easy-to-use system.
|
Number of samples controlled 255_4c1b94-93> | 16 Microcentrifuge Tubes 255_16d9ba-2c> |
|
Designed usage 255_0db57c-07> | Biochemistry 255_2d87c6-7e> |
|
Sample container size 255_c88322-48> | Microcentrifuge Tubes 1.5ml and 2.0ml 255_5602b2-ba> |
|
Optical ports per cuvette 255_0c07bf-d5> | Not Applicable 255_59a04c-c7> |
|
Optical port dimensions 255_fbe058-01> | Not Applicable 255_f673f9-7e> |
|
Built-in optical slit holders 255_9d0e0e-24> | Not Applicable 255_776ed1-9b> |
|
Cuvette z-height 255_b41fda-22> | Not Applicable 255_e16e51-f1> |
|
Windowed jacket or vest available 255_a3aace-fd> | No 255_1c77c7-eb> |
|
Method of heat dissipation 255_b33bda-2a> | Water Circulation 255_55f657-26> |
|
Programable temperature range 255_661bd9-e1> |
-45 to 80°C 255_9adf2e-83> |
|
*Temperature range 255_da25eb-3d> |
-45 to 80°C 255_8dde01-d3> |
|
Extended temperature options available 255_a72f39-77> | No 255_1d709a-a1> |
|
**(Extended) Programable temperature range 255_54d5b7-ee> | Not Applicable 255_7e3636-47> |
|
**(Extended) Temperature range 255_281eff-3c> | Not Applicable 255_d329ef-b5> |
|
Thermistor probes accepted 255_96c2e8-e1> | Probe/10k 255_3eb4a9-e2> |
|
Magnetic stirring motor type 255_85bc34-10> | Not Applicable 255_8d6f08-18> |
|
Default stirring speed 255_17a3b9-d6> | Not Applicable 255_bd35ee-28> |
|
Stirring speed range 255_cdeba6-de> | Not Applicable 255_36dad5-c3> |
|
Stirring speed range (Best performance) 255_8e1c2e-ba> | Not Applicable 255_a94614-a4> |
*The holder's temperature performance depends on the operating environment and setup conditions, which must support proper heat dissipation and condensation reduction. While the listed temperatures are consistently achieved during testing, they are not guaranteed in all environments or setups. Meeting the ideal setup conditions outlined in the product's manual is necessary to achieve specific temperature stages.
**The extended range applies only when all corresponding High or Low customization options are implemented. The higher limit is extended by the High Limit package option, and the lower limit is extended by the Low Limit package option.
Included
Optional
NIST-traceable calibration and performance data

Tubing for dry gas and water circulation

Equipment Cables

TC-1 Temperature Controller

Water Circulator System (BATH 10)

CAL/Single

Probe/10k

T-App Software

NIST-traceable calibration and performance data
Included
Tubing for dry gas and water circulation
Included
Equipment Cables
Included
TC-1 Temperature Controller
Included
Water Circulator System (BATH 10)
Included
CAL/Single
Optional
Probe/10k
Optional
T-App Software
Optional
Manufacture
Model
None
qMC-16